Many automated liquid handlers can detect when a tip touches the liquid surface by measuring a change in capacitance. This is called capacitive liquid-level detection (cLLD or LLD). It lets the instrument follow the meniscus as it aspirates or dispenses, position the tip accurately in Z, and in some cases flag clots or insufficient volume. cLLD only works with conductive tips — ordinary polypropylene is an insulator and is invisible to the sensor.
How Fluotics labels conductive tips
Fluotics encodes conductivity in the SKU with a CB marker, independent of the sterility grade. So a tip can be conductive and filtered/sterile at the same time — for example a …CB.F SKU is a conductive, filtered, sterile tip. Conductive options are available across the lines that need them, including Beckman Biomek, Analytik Jena CyBio, Hamilton, PerkinElmer Janus and Qiagen QIAgility.
When you need conductive tips — and when you don’t
| Your method uses… | Tip type | Why |
|---|---|---|
| Liquid-level detection (cLLD/LLD) | Conductive (CB) | The sensor can only “see” a conductive tip |
| Clot or short-sample detection | Conductive (CB) | Detection relies on the capacitive signal |
| Submerged / following-meniscus aspiration | Conductive (CB) | Z-tracking depends on sensing the surface |
| Fixed-height aspiration, no LLD | Standard | No surface sensing required; standard tips are fine |
Not sure whether your method has LLD enabled? Check the liquid class / pipetting step settings in your instrument software, or ask whoever wrote the method. If LLD is on, use conductive tips; mixing a non-conductive tip into an LLD step will cause errors or aborted aspirations.
→ Browse the full, searchable OEM cross-reference catalog, or request free samples to validate on your own deck.
For the matching cross-reference tables, see the Beckman Biomek, Analytik Jena CyBio, Hamilton STAR/Vantage, PerkinElmer Janus and Qiagen QIAgility guides, or the complete cross-reference index.
Frequently asked questions
What does CB mean on an automation pipette tip?
CB denotes a conductive tip. Conductive tips are made from a carbon-loaded polymer so the instrument can detect the liquid surface capacitively (liquid-level detection, or cLLD).
Which instruments need conductive tips?
Liquid handlers that use capacitive liquid-level detection — for example many Hamilton Microlab STAR/Vantage, Beckman Biomek, Analytik Jena CyBio and PerkinElmer Janus configurations. Check whether your method has LLD enabled; if it does, you need conductive tips.
Do I always need conductive tips?
No. If your protocol uses fixed-height aspiration (no liquid-level detection), standard non-conductive tips are fine and usually less expensive. Conductive tips are required when the instrument relies on cLLD for accurate Z-positioning, clot detection or low-volume work.
Are conductive tips available filtered and sterile?
Yes — Fluotics offers conductive tips in non-sterile (.NS), sterile (.ST) and filtered/sterile (.F) grades, encoded in the SKU (e.g. a CB tip can also be .F).
Can I get free samples of conductive tips?
Yes — request free samples to confirm liquid-level detection performs correctly with your method before ordering a case.
Instrument and product names referenced are trademarks of their respective owners (Agilent, Tecan, Beckman Coulter, Analytik Jena, Revvity/PerkinElmer, Hamilton, QIAGEN, Molecular Devices). Fluotics is an independent manufacturer and is not affiliated with or endorsed by these companies; OEM part numbers are listed solely for cross-reference and compatibility identification.
Related Fluotics pages & guides
- Complete automation pipette-tip OEM cross-reference guide
- Choosing 96- vs 384-format automation tips
- How Fluotics tips are validated against OEM performance
- Molecular Devices FLIPR compatible-tip cross-reference
- Beckman Biomek compatible-tip cross-reference
- Browse the full Fluotics compatible-tip catalog